| Model Number |
180-01-50 |
| Medical Device Problem Code |
Loosening of Implant Not Related to Bone-Ingrowth (4002)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
09/11/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Pending engineering evaluation.Concomitant medical products: cocr femoral head 32mm +3.5 offset (cn: 142-32-03, sn: (b)(4); nv gxl liner 10o 32mm (cn: 138-32-51, sn: (b)(4); screw 30mm 6.5 (cn: 120-65-30, sn: (b)(4); c-series 2nd offset 12/14 size 1 (cn: 116-30-01, sn: (b)(4).
|
| |
|
Event or Problem Description
|
|
Original right total hip arthroplasty was performed on this female patient on (b)(6) 2009.The surgeon brought the patient in because their acetabular cup and liner were spun out and loose when he looked at their x-ray.The surgeon opened the joint in standard fashion.The surgeon decided to pull everything out and implant stryker femoral components and acetabular components.The patient left the operating room in stable condition and is expected to have a good outcome.There was no delay to the surgery.The rep was present at the time of the surgery.The hospital did not release implants to be returned to exactech.
|
| |
|
Additional Manufacturer Narrative
|
|
The evaluation noted that the reason for the revision reported was likely the result of an insufficient bond between the implant and the bone, which led to aseptic (non-infected) loosening of the acetabular cup.However, this cannot be confirmed since the devices were not returned for evaluation.(d11) concomitant device(s): cocr femoral head 32mm +3.5 offset (cn:(b)(4), sn: (b)(6).Nv gxl liner 10o 32mm (cn:(b)(4), sn: (b)(6).Screw 30mm 6.5 (cn:(b)(4), sn: (b)(6).C-series 2nd offset 12/14 size 1 (cn:(b)(4), sn: (b)(6).
|
| |
|
Search Alerts/Recalls
|