MAUDE Adverse Event Report: NEURO - VILLALBA VECTRIS SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NEURO - VILLALBA VECTRIS SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) Back to Search Results
Model Number 977A260
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from a healthcare provider (hcp) via a manufacturer representative (rep) regarding a patient with an implantable neurostimulator (ins).It was reported that the patient didn't like having a rechargeable battery so the decision was made to replace it with a non-rechargeable battery.During the replacement, intra-op impedance testing showed all impedances were open.The hcp opened another new ins and again all impedances were open.The leads were tested with a wens and all impedances were open.The leads were replaced with two new leads and all impedances were then normal.The rep didn't know what led to the issue.The issue was reported to be resolved.The explanted leads were expected to be returned for analysis.No further complications were reported.
 
Additional Manufacturer Narrative
H3: analysis of the lead (s/n: (b)(6) found all conductors were broken 19.8 cm from the proximal end.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VECTRIS SURESCAN
Common Device Name
STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
Manufacturer (Section D)
NEURO - VILLALBA
call box 6001
villalba PR 00766
MDR Report Key9139925
Report Number6000153-2019-00012
Device Sequence Number3046230
Product Code LGW
UDI-Device Identifier00643169109391
UDI-Public00643169109391
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
P840001
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2019
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 02/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/18/2019
Device Model Number977A260
Device Catalogue Number977A260
Device Lot NumberVA0WZNF037
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/08/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2019
Supplement Date Received by Manufacturer02/06/2020
Initial Report FDA Received Date10/01/2019
Supplement Report FDA Received Date02/11/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-