| Model Number |
977A260 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from a healthcare provider (hcp) via a manufacturer representative (rep) regarding a patient with an implantable neurostimulator (ins).It was reported that the patient didn't like having a rechargeable battery so the decision was made to replace it with a non-rechargeable battery.During the replacement, intra-op impedance testing showed all impedances were open.The hcp opened another new ins and again all impedances were open.The leads were tested with a wens and all impedances were open.The leads were replaced with two new leads and all impedances were then normal.The rep didn't know what led to the issue.The issue was reported to be resolved.The explanted leads were expected to be returned for analysis.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: analysis of the lead (s/n: (b)(6) found all conductors were broken 19.8 cm from the proximal end.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
| |
|
Search Alerts/Recalls
|