Investigation summary: three sample were received.They have the plunger rod-rubber stopper, only one has tip cap and all have barrel label; they have no packaging flow wrap.The plunger rod-rubber stopper is at 3.5ml, 5ml, and the third sample at 6ml.All three have solution.Visual inspection was performed, none of them has barrel deformation or any other type of damage.The sample with tip cap was tested for sustaining force.The other two samples with no tip cap were checked for silicone dispersion finding them with poor silicone distribution at the bottom part of the syringe.Possible root cause, barrel silicone application.We can confirm the complaint from the fact that even when one sample was within spec for sustaining force, this was at the high side of the product specification; the other two were tested for silicone dispersion to understand why the first sample was on the high side of the specification.The silicone dispersion was satisfactory.Note: after performing the sustain force we cannot perform the silicone dispersion, since the rubber stopper would place some silicone on the inner barrel wall as it goes down during the test.No corrective action at this time.Complaints received for this device and reported condition will continue to be tracked and trended.Information will be captured on trend reports and monitored monthly.Our business team regularly reviews the collected data for identification of emerging trends.
|
|
It was reported that bd posiflush¿ surescrub pre-filled syringe would cease up.This occurred on 3 occasions during use.The following information was provided by the initial reporter: material no.: 306559, batch no.: 9136939.It was reported that the pre-filled syringe would cease up because the stopper would stick and the barrel of the syringe was tapered.Pr 1 of 2: this pr is for incident that occurred "and the same issue arose again today 3 times." we have 52 boxes of 30 each with the same lot.We just received 2 cases on it today.We had an isolated issue last week and the same issue arose again today 3 times.2 were back to back.When the iv was started and the nurses were going to flush, the prefilled syringes would cease up.The stopper would stick around the 7-8mm spot.It is almost like the syringe was tapered not to allow the plunger to advance any further.We only have 1 box of 30 on our shelf.
|