| Model Number |
24673 |
| Medical Device Problem Code |
Difficult to Insert (1316)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age at time of event: 18 years or older.Date of event: used (b)(6) 2019 as no event date was provided.Device evaluated by mfr.: a visual and microscopic examination was performed on the returned mustang device.A visual examination of the returned device confirmed that there was a longitudinal tear observed in the balloon material.The longitudinal tear in the balloon stretches for 100mm from the proximal edge of proximal markerband to the middle section of the balloon material.During analysis the investigator did not attempt to advance the device through a 5fr boston scientific sheath due to the condition of the balloon material.A visual and microscopic examination found no damage or any issues with the tip or markerbands of the device which could have contributed to the complaint incident.A visual and tactile examination found no kinks or damage along the shaft of the device.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 12sep2019.It was reported that insertion difficulty was encountered.A 6.0 x 200, 75cm mustang balloon catheter was selected for use, but they encountered difficulty in going through the non-bsc 5fr sheath.There were no patient complications reported.However, returned device analysis revealed a longitudinal tear in the balloon.
|
| |
|
Search Alerts/Recalls
|