| Medical Device Problem Codes |
Image Display Error/Artifact (1304); Structural Problem (2506)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation: the pump has been returned and evaluated by product analysis on 30-jul-2019 with the following findings: during investigation, the battery compartment was found to be cracked.Further investigation found the display to be dim and discolored.Animas has conducted a review of the device history record for this pump and confirmed that it was operating within required specifications at the time of release.Report late due to transition from vmsr program.(b)(6).
|
| |
|
Event or Problem Description
|
|
The pump was returned for investigation.Investigation revealed a cracked battery compartment.Further investigation also found that the display was dim and discolored.This report is made based on results of investigation completed on 30-jul-2019.There was no indication that the product caused or contributed to an adverse event.
|
| |
|