• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM A. COOK AUSTRALIA, PTY LTD ZENITH FENESTRATED AAA ENDOVASCULAR GRAFT PROXIMAL BODY; MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM A. COOK AUSTRALIA, PTY LTD ZENITH FENESTRATED AAA ENDOVASCULAR GRAFT PROXIMAL BODY; MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number G32546
Medical Device Problem Code Appropriate Device Problem Term/Code Not Available (3191)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 05/29/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
This event has been reported by the manufacturer under mdr# 3002808486-2019-00032.
 
Event or Problem Description
Following the procedure for implantation, there was a leak at the distal edge of the overlap between the two devices on completion angiogram.The source of the leak could not be determined.The physician implanted two additional covered stents extending slightly above the flow divider and post dilated them.Another angiogram following this showed the leak to be repaired but the exact source of the leak was not identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZENITH FENESTRATED AAA ENDOVASCULAR GRAFT PROXIMAL BODY
Common Device Name
MIH SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
WILLIAM A. COOK AUSTRALIA, PTY LTD
95 brandl street
eight mile plains
brisbane QLD 4 113
AS  QLD 4113
MDR Report Key9141851
Report Number3005580113-2019-00447
Device Sequence Number3185739
Product Code MIH
UDI-Device Identifier10827002325460
UDI-Public(01)10827002325460(17)220426(10)AC1036278
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/01/2019
Report Date (Section F) 10/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date04/26/2022
Device Model NumberG32546
Device Catalogue NumberZFEN-P-2-36-107-R
Device Lot NumberAC1036278
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date09/10/2019
Event Location Hospital
Date Report to Manufacturer10/01/2019
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date10/01/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age64 YR
Patient Weight80
Date Report Sent to FDA10/01/2019
-
-