• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK IRELAND LTD ECHOTIP PROCORE ENDOBRONCHIAL HD BIOPSY NEEDLE; FCG KIT, NEEDLE, BIOPSY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD ECHOTIP PROCORE ENDOBRONCHIAL HD BIOPSY NEEDLE; FCG KIT, NEEDLE, BIOPSY Back to Search Results
Model Number G34279
Device Problem Break (1069)
Patient Problem Foreign Body In Patient (2687)
Event Date 09/03/2019
Event Type  Injury  
Manufacturer Narrative
Pma/510(k) #k160229.1 unit of lot number c1609874 of echo-hd-22-ebus-p-c was returned opened not in its original packaging.The device related to this occurrence underwent a laboratory evaluation.The needle was found to be broken distally.Prior to distribution, all echo-hd-22-ebus-p-c devices are subjected to functional checks and visual inspection to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place at cook (b)(4).A review of the manufacturing records for echo-hd-22-ebus-p-c of lot number c1609874 did not reveal any discrepancies that could have contributed to this complaint issue.The review of relevant manufacturing records, confirms the failure mode has not previously occurred with the current lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1609874.The notes section of the instructions for use, which accompanies this device instructs the user to; "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use".There is no evidence to suggest that the customer did not follow the instructions for use.The failure of needle broken was concluded from the available information.A definitive root cause for the customer complaint could not be determined as circumstances of use cannot be replicated in the laboratory.A possible root cause could be attributed to a combination of tortuous position, flexed or twisted endoscope position and hard lesion as indicated in the additional information.Complaint is confirmed as the failure was verified in the laboratory and in the images shared.The tip of the needle has been retrieved from the patient with a forceps.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event Description
Needle got broken while taking sample from the subcranial region approximately 1.5 cm from the tip.The tip got detached and retrieval through flexible scope using forceps.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHOTIP PROCORE ENDOBRONCHIAL HD BIOPSY NEEDLE
Type of Device
FCG KIT, NEEDLE, BIOPSY
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
Manufacturer (Section G)
COOK IRELAND LTD
o halloran road
national technology park
limerick
Manufacturer Contact
heather ryan
o halloran road
national technology park
limerick 
061334440
MDR Report Key9144140
MDR Text Key167504975
Report Number3001845648-2019-00500
Device Sequence Number1
Product Code FCG
UDI-Device Identifier00827002342798
UDI-Public(01)00827002342798(17)220501(10)C1609874
Combination Product (y/n)N
Reporter Country CodeIN
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 09/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/01/2022
Device Model NumberG34279
Device Catalogue NumberECHO-HD-22-EBUS-P-C
Device Lot NumberC1609874
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/24/2019
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date09/03/2019
Event Location Hospital
Initial Date Manufacturer Received 09/04/2019
Initial Date FDA Received10/02/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age60 YR
-
-