• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. TRICUT® STRAIGHT SHAFT 4MM 11CM LONG; BUR, EAR, NOSE AND THROAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. TRICUT® STRAIGHT SHAFT 4MM 11CM LONG; BUR, EAR, NOSE AND THROAT Back to Search Results
Model Number 1884004
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/10/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A health care provider (hcp) reported that a blade was fractured during the procedure and was completed using a back up product.No broken pieces of the reported product remain inside the patient's body.There was no patient impact.
 
Additional Manufacturer Narrative
H3: analysis found that visually, the distal tip of the inner cutter broke off which would have resulted in the reported event.The breakage occurred at the first proximal tooth valley.The portion that broke off was not returned however it would have measured approximately 0.24¿ in length.Note ¿ the distal tip outside diameter of the inner cutter (expanded area) did not appear to have a turned / machined finish when compared the remainder of the tube.The outer assembly was deformed in such a way that indicates the inner blade teeth impacted the outer teeth at the 4th proximal tooth while moving in a cw direction and the 3rd proximal valley while moving in a ccw direction which resulted in the observed break.There were no signs of misuse, mishandling bends, or concentricity issues.There was uneven wear at the cutting tip which may indicate an interference fit between the inner and outer assemblies at the tip.The inner cutter and outer tube are supplier manufactured; additional components are installed for the final assembled by medtronic at the jacksonville and juncos manufacturing sites.H6: fdm - 4114 ; fdr - 3221 ; fdc - 67 are no longer applicable.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRICUT® STRAIGHT SHAFT 4MM 11CM LONG
Common Device Name
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
MDR Report Key9144862
Report Number3004209178-2019-18793
Device Sequence Number16853517
Product Code EQJ
UDI-Device Identifier00681490047456
UDI-Public00681490047456
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1884004
Device Catalogue Number1884004
Device Lot NumberHG2Z6HF
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2019
Supplement Date Received by Manufacturer03/23/2020
Initial Report FDA Received Date10/02/2019
Supplement Report FDA Received Date04/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-