| Model Number |
CTRF-220 |
| Medical Device Problem Codes |
Break (1069); Overheating of Device (1437); Smoking (1585); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Code Available (3191)
|
| Date of Event |
09/03/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during procedure, after the physician place the cluster electrode in the patient body and once the machine was switched on, the machine started to heat up suddenly and there were burning smoke sensation occurred.The procedure was stopped and did not proceed with ablating.Whole set of machines and cluster electrode were return back to the office.
|
| |
|
Event or Problem Description
|
|
According to the reporter, during procedure, after the physician place the cluster electrode in the patient body and once the machine was switched on, the machine started to heat up suddenly and there were burning smoke sensation occurred.The procedure was stopped and did not proceed ablating using radio frequency ablation.Surgeon switched to emprint antenna to complete the procedure with no post-procedural adverse event happened to the patient.Whole set of machines and cluster electrode were return back to the office.Service engineer at the facility found out the problem that caused the smoke from the unit was a faulty system main circuit board.
|
| |
|
Additional Manufacturer Narrative
|
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|