Catalog Number 1012445-08
Medical Device Problem Codes Material Separation (1562); Deformation Due to Compressive Stress (2889)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/19/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was returned for analysis with the hypotube separated.The reported kink was able to be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that inadvertent mishandling during preparation for use and/or advancement onto the guide wire resulted in the reported kink, the noted multiple bends and ultimately resulted in the noted shaft separation.There is no indication of a product quality issue with respect to manufacture, design or labeling.The 3.5 x 12 mm nc trek device referenced is being filed under a separate medwatch report number.Exemption number e2019001.
 
Event or Problem Description
It was reported that during preparation, the shafts of a 2.0 x 8 mm and a 3.5 x 12 mm nc trek balloon catheter became kinked.Therefore, a non-abbott balloon catheter was used to successfully complete the procedure.There was no patient involvement.No additional information was provided.Returned device analysis revealed that both devices were used and that the hypotubes were separated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK RX
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9145789
Report Number2024168-2019-12287
Device Sequence Number14119461
Product Code LOX
UDI-Device Identifier08717648151637
UDI-Public08717648151637
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2022
Device Catalogue Number1012445-08
Device Lot Number90422G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/04/2019
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/11/2019
Initial Report FDA Received Date10/02/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-