Model Number 14500-5A
Device Problem Unable to Obtain Readings (1516)
Patient Problem Hyperglycemia (1905)
Event Date 05/07/2019
Event Type  malfunction  
Manufacturer Narrative
The pdm was found to have a nonfunctioning bg meter board, resulting in a meter error 4.The pdm would not give bg readings upon strip insertions or sst test.Replacing the bg board resolved this issue.Lot release records were reviewed and the product lot met all acceptance criteria.Omnipod insulin management system ¿ user guide: model: ust400, 17845-5a-aw rev b 09/17.Checking your blood glucose: chapter 4 / page 36; you should perform a control solution test when: you suspect that the built-in bg meter or test strips are not working properly.You think your bg readings are not accurate or are not consistent with how you feel.You drop or damage your pdm or expose it to liquids.Your healthcare provider advises you to do so.When you perform a control solution test, if the reading is within the control solution acceptable range, the built-in bg meter is working properly.With the built-in bg meter, checking your blood glucose requires a very small sample size, 0.3 microliters of blood.Checking your blood glucose : chapter 4 / page 43; warning: "low" or "high" blood glucose readings can indicate a potentially serious condition requiring immediate medical attention.If left untreated, these situations can quickly lead to diabetic ketoacidosis (dka), shock, coma, or death.Consult your healthcare provider for how to treat high and low blood glucose levels.Living with diabetes: chapter 11 / page 116; warning: keep an emergency kit with you at all times to quickly respond to any diabetes emergency.Prepare an emergency kit to keep with you at all time.The kit should include: several new, sealed pods.Extra new pdm batteries (at least two aaa alkaline; do not use rechargeable batteries).A vial of rapid-acting u-100 insulin (see the introduction for insulins approved for use in the omnipod® system).Syringes or pens for injecting insulin.Blood glucose test strips.Additional blood glucose meter.Ketone test strips.Lancing device and lancets.Glucose tablets or another fast-acting source of carbohydrate.Alcohol prep swabs.Instructions from your healthcare provider about how much insulin to inject if delivery from the pod is interrupted.A signed letter from your healthcare provider explaining you need to carry insulin supplies and omnipod® system equipment.Phone numbers for your healthcare provider and/or physician in case of an emergency.Glucagon kit and written instructions for giving an injection if you are unconscious (see "avoid lows, highs, and dka" on page 119).Living with diabetes: chapter 11 / page 119; avoid lows, highs, and dka: you can avoid most risks related to using the omnipod® system by practicing proper techniques and by acting promptly at the first sign of hypoglycemia, hyperglycemia, or diabetic ketoacidosis.The easiest and most reliable way to avoid these conditions is to check your blood glucose often.
 
Event Description
It was reported that the patient's blood glucose (bg) levels reached 432 mg/dl.The pdm (personal diabetes manager) was displaying meter error 4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNIPOD INSULIN PUMP
Type of Device
PUMP, INFUSION, INSULIN
Manufacturer (Section D)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer (Section G)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer Contact
jordan biswurm
100 nagog park
acton, MA 01720
9786007000
MDR Report Key9146179
MDR Text Key190773629
Report Number3004464228-2019-09575
Device Sequence Number1
Product Code LZG
UDI-Device Identifier10385081140003
UDI-Public(01)10385081140003(11)170811(10)L60719
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122953
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Other
Type of Report Initial
Report Date 05/07/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/02/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number14500-5A
Device Catalogue NumberUST400
Device Lot NumberL60719
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2019
Date Manufacturer Received09/14/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/11/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age7 YR
-
-