Concomitant medical products: unk balloon catheter, unk stent.Complaint event summary: the power flex pro (4mm x 4cm x 80cm) was used for the iliac artery which had heavy calcification.However, it ruptured at ten atmospheres (atm) during its initial inflation.The procedure was continued with another non-cordis balloon catheter and a stent was implanted.There was no reported patient injury.The product was not returned for analysis.A product history record (phr) review of lot 17578603 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿balloon burst-at/below rbp¿ could not be confirmed as the device was not returned for analysis.The exact cause could not be determined.With the paucity of information available and without return of the product for analysis, it is not possible to draw a clinical conclusion between the device and the reported event.However, vessel characteristics and procedural factors may have contributed to the reported event.According to the safety information in the instructions for use ¿if resistance is met during manipulation, determine the cause of the resistance before proceeding.Balloon pressure should not exceed the rated burst pressure.The rated burst pressure is based on the results of in vitro testing.At least 99.9% of the balloons (with a 95% confidence), will not burst at or below their rated burst pressure.Use of a pressure monitoring device is recommended to prevent over pressurization.Use only the recommended balloon inflation medium.Never use air or any gaseous medium to inflate the balloon.Prior to angioplasty, the catheter should be examined to verify functionality and ensure that its size and shape are suitable for the specific procedure for which it is to be used.¿ neither the phr nor the information available suggests a design or manufacturing related cause for the reported event.Therefore, no corrective or preventive action will be taken at this time.
|
|
As reported, the power flex pro (4mm4cm 80) was used for the iliac artery which had heavy calcification.However, it ruptured at 10 atmospheres (atm) during its initial inflation.The procedure was continued with another non-cordis balloon catheter and a stent was implanted.There was no reported patient injury.The device will not be returned for analysis.
|