• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL DELTAPAQ CERE 1.5MMX2CM; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL DELTAPAQ CERE 1.5MMX2CM; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Catalog Number CDF10015230
Medical Device Problem Codes Stretched (1601); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 08/03/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #: (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Procode: krd/hcg.Initial reporter: the customer contact information, including name, occupation, phone, fax, and e-mail address, was not reported.Manufacturing site name: codman and shurtleff inc., dba depuy synthes products, inc.(b)(4).The device is available to be returned for evaluation and testing.However, it has not been received to date.If the device returns, a device investigation will be performed.The company is seeking this information through the event investigation.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.
 
Event or Problem Description
As reported by a healthcare professional, during a coil embolization, a 1.5mmx2cm deltapaq cere coil (cdf10015430, s14242) was being used as the first coil, however, the coil couldn't be advanced nor retrieved by the physician.The physician pulled the coil and the unknown microcatheter at the same time.After inspection it turns out to be stretched.The physician used a same like product.The procedure was successfully finished.There was no report of patient injury.
 
Additional Manufacturer Narrative
Product complaint#: (b)(4).Updated sections on this report: d10 (concomitant medical products), g4 (pma/510k), h2 (if follow-up, what type), h3 (device evaluated by mfr), h6 and h10.Complaint conclusion: as reported by a healthcare professional, during a coil embolization, a 1.5mmx2cm deltapaq cere coil (cdf10015430, s14242) was being used as the first coil, however, the coil couldn't be advanced nor retrieved by the physician.The physician pulled the coil and the unknown microcatheter at the same time.After inspection it turns out to be stretched.The physician used a same like product.The procedure was successfully finished.There was no report of patient injury.One non-sterile unit deltapaq cere 1.5mmx2cm was received inside of a pouch.The hub, the dpu and the introducer were inspected, and no damages were noted on them.The re-sheathing tool was inspected and it was found in good conditions.The v-notch was inspected under microscope and no damages were noted on it.The rh was inspected under microscope and it was found in good conditions.The embolic coil was inspected under microscope and no damages were found on it.A lab sample prowler select plus was flushing using a lab sample syringe.After that, the unit deltapaq cere 1.5mmx2cm was introduced in to a lab sample prowler select plus and it advance, no resistance/friction was felt.A manufacturing record evaluation was performed for the finished device s14242 number, and no non-conformances related to the reported complaint condition were identified.The complaint reported by the customer ¿detachable coil delivery system (dcs) - impeded in microcatheter with loss of cerebral target position¿ was not confirmed due to in the functional test the received device could pass through the mc.The complaint reported by the customer ¿coil - unraveled/stretched-in microcatheter¿ was not confirmed due to no damages were found on the embolic coil.Neither the analysis nor the mre suggest that the failure reported by the costumer could be related to the manufacturing process.Assignment of root cause for the events remains speculative and inconclusive, based on the limited information provided.However, it is possible that clinical and procedural factors, including device manipulation / interaction, aneurysm size and vessel characteristics, device selection, and the concomitant microcatheter, may have contributed to the reported issues.Assignment of root cause for the event remains speculative and inconclusive, based on the limited information provided and without the return of the complaint devices; however, it is possible that clinical and procedural factors, including device manipulation / interaction, aneurysm size and vessel characteristics, device selection, and the concomitant microcatheter, may have contributed to the reported issue.Multiple attempts to obtain additional information were unsuccessful.The instructions for use (ifu) warns that if the microcoil system becomes immobile in the infusion microcatheter, apply a gentle push-pull motion to free it.If unsuccessful, remove both microcatheter and microcoil system together as a unit and replace with new devices.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTAPAQ CERE 1.5MMX2CM
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
MDR Report Key9147661
Report Number3008114965-2019-01192
Device Sequence Number2664945
Product Code HCG
UDI-Device Identifier00878528007012
UDI-Public00878528007012
Combination Product (Y/N)N
Initial Reporter CountryKS
PMA/510(K) Number
K080437
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2022
Device Catalogue NumberCDF10015230
Device Lot NumberS14242
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2019
Initial Date Received by Manufacturer 09/17/2019
Supplement Date Received by Manufacturer10/14/2019
Initial Report FDA Received Date10/02/2019
Supplement Report FDA Received Date10/31/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN MICROCATHETER
-
-