• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG., S.J. CUTTER, AGG. MENICUS - 2.5MM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG., S.J. CUTTER, AGG. MENICUS - 2.5MM; BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL Back to Search Results
Catalog Number 0275628000
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/03/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.
 
Event or Problem Description
It was reported that the tip of the device broke during the procedure.No pieces fell into the joint space.
 
Event or Problem Description
It was reported that the tip of the device broke during the procedure.No pieces fell into the joint space.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: inner blade metal tip broken off.Probable root cause: inadequate window profile, inadequate welding process (i.E.Plasma, inductive, gluing), improper material strength, inadequate size selected for the application, material brittleness, excessive force applied by the user, incorrect rfid tag.The reported failure mode will be monitored for future reoccurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG., S.J. CUTTER, AGG. MENICUS - 2.5MM
Common Device Name
BLADE, SAW, GENERAL & PLASTIC SURGERY, SURGICAL
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key9148026
Report Number0002936485-2019-00444
Device Sequence Number2132197
Product Code GFA
UDI-Device Identifier37613327061568
UDI-Public37613327061568
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 12/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number0275628000
Device Lot Number19017CG2
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/09/2019
Supplement Date Received by Manufacturer09/09/2019
Initial Report FDA Received Date10/02/2019
Supplement Report FDA Received Date12/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-