| Model Number |
MMT-751NAH |
| Medical Device Problem Code |
Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/30/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
The customer's family member reported via phone call that the customer's insulin pump had button anomaly.The customer¿s blood glucose was 219 mg/dl at the time of incident.The customer stated that buttons not responding specific to act button and the up button was also starting to be sticking.The customer stated that the time was advancing.The customer was advice the insulin pump will need to be replaced.Advice to discontinue use of the insulin pump and revert to backup plan per health care professional instructions.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with intermittent button response due to flattened (escape and act ) button dome switch (no crease) and partial unlocked lcd keypad connector noted during visual inspect.Unable to perform all functional testing including the displacement, operating currents, a21 error test, rewind, basic occlusion, occlusion, prime or a33 and excessive no delivery test due to buttons not responding.
|
| |
|