|
(b)(4).The device was not returned to the service center.As part of our investigation, the service history for the subject device and found that the device was evaluated on april 18, 2019 by the manufacturer¿s clinical specialist and the image issue was confirmed.There was no record of service for the subject device.The investigation is currently ongoing to obtain additional information regarding the reported event.
|
|
This supplemental report is being submitted to report the device evaluation results.Please the updates in sections: d10, g4, g7, h2, h3, h6 and h10.An endoscopic ultrasound eu-me2 premier plus, serial# (b)(6), was returned for evaluation also, based on previous service order 19075247 ¿description of the problem ¿ the near field is very dark and the progressive is very bright ¿ there are 3 lines on contrast¿.The device came with a new keyboard maj-1995, sn (b)(6).A visual inspection was performed on the as received condition of the device and found the front panel with a large crack on its lower right corner.The rear panel was also bent at the ac inlet area.There are 2 input sockets on the front; however, the metal frame of the socket #1 was damaged; therefore, the socket 1 could not be tested for functionality.The socket #2 was then checked with our test us cable maj-1597 and an ultrasound endoscope bf-uc180f.The us image displayed on the monitor, but corrupted.It was very bright, and not dark or grainy as reported.Based on the evaluation findings, the reported complaint of bright image was confirmed.However, dark and grainy image was not duplicated.The physical damage found on the front panel and sockets could have contributed to the reported complaint.This damage is likely due to mishandling as the device could have been dropped causing a strong impact to the device.
|