Model Number 1103
Device Problems Device Damaged Prior to Use (2284); Material Integrity Problem (2978)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/23/2019
Event Type  malfunction  
Manufacturer Narrative
This device is used for treatment not diagnosis.The ventricular assist system is indicated for use as a bridge to cardiac transplantation and destination therapy in patients who are at risk of death from refractory end-stage left ventricular heart failure.The system is designed for in-hospital and out-of-hospital settings.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during the implant procedure, the o-ring on the ventricular assist device (vad) was damaged out of the sterile packaging.The vad was not used and a different vad was implanted and remains in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the ventricular assist device (vad) and the o-ring were returned for evaluation.Review of the manufacturing documentation confirmed that the pump and associated o-ring met all requirements for release.Various analyses were conducted and reviewed in order to evaluate the performance of the returned devices in relation to the reported event.Failure analysis of the returned pump revealed that the device passed external visual inspection.Failure analysis of the returned o-ring revealed a cut, most likely introduced during the installation process.As a result, the reported "damaged o-ring" event was confirmed.There is no evidence to suggest that a device malfunction caused or contributed to the reported event.An internal investigation evaluated o-ring damages.Based on that investigation, the most likely root cause of the reported event can be attributed to multiple factors such as possible variation in the sewing ring gap that may result in variation in the effective diameter of the sewing ring as assembled and the angle of insertion by physician during implant.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTWARE VENTRICULAR ASSIST SYSTEM - PUMP/DRIVELINE
Type of Device
VENTRICULAR (ASSIST) BYPASS
Manufacturer (Section D)
HEARTWARE, INC.
14400 nw 60th ave
miami lakes FL 33014
MDR Report Key9151527
MDR Text Key197775378
Report Number3007042319-2019-12014
Device Sequence Number1
Product Code DSQ
UDI-Device Identifier00888707003261
UDI-Public00888707003261
Combination Product (y/n)N
PMA/PMN Number
P100047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/31/2021
Device Model Number1103
Device Catalogue Number1103
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2019
Initial Date Manufacturer Received 09/23/2019
Initial Date FDA Received10/03/2019
Supplement Dates Manufacturer Received11/15/2019
Supplement Dates FDA Received11/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age58 YR
-
-