| Catalog Number |
217863174 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
Not Applicable (3189)
|
| Date of Event |
09/12/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary = > examination of the returned device confirmed the reported event.The investigation attributed the root cause to normal use and servicing and did not establish that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the surgeon was reaming with the hudson adaptor and reamers the hudson attachment and the reamers stripped and had difficulty reaming the femur and the tibia.They tried this with both sets of our hudson¿s and t-handles and reamers.Also the size 4 tc3 box trial would not stay on the femur trial so think the torque screw must be stripped.A surgical delay of 15-20 minutes.
|
| |
|
Search Alerts/Recalls
|