• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: VYAIRE MEDICAL MICROLAB; SPIROMETER, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

VYAIRE MEDICAL MICROLAB; SPIROMETER, DIAGNOSTIC Back to Search Results
Model Number 36-ML3535 MK8-STK
Medical Device Problem Code Smoking (1585)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/10/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Vyaire file identification: (b)(4).Any additional information received from the customer will be included in a follow-up report.The customer reported the burning smell came from the electronics board.At this time, vyaire has not received the suspected component for evaluation.
 
Event or Problem Description
The customer reported burnt electronic board from the microlab-loop device.The customer reported the equipment began to smell like smoke and was identified by the technician through the burning smell.The customer reported there is no patient involvement associated with the event.
 
Additional Manufacturer Narrative
Results of investigation: the vyaire failure analysis lab received the suspected device/component and performed a failure investigation.The reported issue could not be evaluated in the laboratory setting due to the burnt component, making it impossible to evaluate.The device was repaired and the unit has passed all test, calibrations and is working to manufacture specifications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICROLAB
Common Device Name
SPIROMETER, DIAGNOSTIC
Manufacturer (Section D)
VYAIRE MEDICAL
22745 savi ranch pkwy
yorba linda CA 92887
MDR Report Key9151983
Report Number9615102-2019-00110
Device Sequence Number2131874
Product Code BZG
UDI-Device Identifier54250892905469
UDI-Public(01)54250892905469(11)20170105
Combination Product (Y/N)N
Initial Reporter CountryPO
PMA/510(K) Number
K141936
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 09/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number36-ML3535 MK8-STK
Device Catalogue Number36-ML3500 MK8-STK
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 09/10/2019
Supplement Date Received by Manufacturer09/26/2019
Initial Report FDA Received Date10/03/2019
Supplement Report FDA Received Date11/06/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-