| Model Number |
30935 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/01/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event is an estimated date based off the aware date as the exact event date was not reported.Device evaluated by manufacturer: the device was returned for analysis.The catheter returned has the hub detached (the heat shrink was received properly attached to the hub); moreover, the flare of the extrusion was found formed, it is evidence that the manufacturing process was properly performed.The catheter returned with detached hub.The catheter was also cut at 12cm from the flare end.The rest of the catheter was not returned for analysis.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 12sep2019.A flexima apdl catheter drainage was received with no reported issues.However, analysis revealed hub detached.
|
| |
|