Model Number 30935
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/01/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event is an estimated date based off the aware date as the exact event date was not reported.Device evaluated by manufacturer: the device was returned for analysis.The catheter returned has the hub detached (the heat shrink was received properly attached to the hub); moreover, the flare of the extrusion was found formed, it is evidence that the manufacturing process was properly performed.The catheter returned with detached hub.The catheter was also cut at 12cm from the flare end.The rest of the catheter was not returned for analysis.
 
Event or Problem Description
Reportable based on device analysis completed on 12sep2019.A flexima apdl catheter drainage was received with no reported issues.However, analysis revealed hub detached.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXIMA APDL
Common Device Name
TUBE, DRAINAGE, SUPRAPUBIC
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key9152291
Report Number2134265-2019-11948
Device Sequence Number2782643
Product Code FFA
UDI-Device Identifier08714729323143
UDI-Public08714729323143
Combination Product (Y/N)N
Initial Reporter StateWI
Initial Reporter CountryUS
PMA/510(K) Number
K944290
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/15/2022
Device Model Number30935
Device Catalogue Number30935
Device Lot Number0023803685
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2019
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/12/2019
Initial Report FDA Received Date10/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-