• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR NC TREK RX; CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR NC TREK RX; CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012455-20
Medical Device Problem Codes Off-Label Use (1494); Difficult to Remove (1528); Material Separation (1562)
Health Effect - Clinical Code Embolism (1829)
Date of Event 09/12/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was received.Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.Exemption number e2019001-permits numbering sequence to begin with 10000, to avoid duplication of report numbers due to process transition.There may be gaps in numbering for reports submitted during the transition period.
 
Event or Problem Description
It was reported that the patient presented with a heavily tortuous iliac artery with heavy calcification.The nc trek dilatation catheter advanced and dilatation was performed one time at 14 atmospheres without issues.The balloon was deflated for withdrawal.Upon retraction into the sheath, unusual resistance was met and the balloon had separated from the delivery catheter and embolized into the aorta.A snare successfully removed the separated portion.Nothing was left in the patient's anatomy.There was no adverse patient sequela and no clinically significant delay reported.
 
Additional Manufacturer Narrative
Visual inspection was performed on the returned device.The reported separation was confirmed.The reported difficulty removing the device could not be replicated in a testing environment due to the condition of the returned device.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents and/or complaints from this lot.The reported patient effect of embolism is listed in the coronary dilatation catheters (cdc), nc trek rx, global, instruction for use as a known patient effect.It was reported that the device was used to treat the iliac artery.It should be noted that the coronary dilatation catheters (cdc), nc trek rx instructions for use states: the nc trek rx coronary dilatation catheters are indicated for: balloon dilatation of the stenotic portion of a coronary artery or bypass graft stenosis, for the purpose of improving myocardial perfusion and balloon dilatation of a coronary artery occlusion, for the purpose of restoring coronary flow in patients with st-segment elevation myocardial infarction.The investigation was unable to determine a conclusive cause for the reported difficulty removing the device; however, the reported separation, removal of foreign body and additional treatment appear to be related to operational context.A conclusive cause for the reported patient effect of embolism and the relationship to the device, if any, cannot be determined.There is no indication of a product quality issue with respects to the design, manufacture, or labeling of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK RX
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key9152412
Report Number2024168-2019-12333
Device Sequence Number2547295
Product Code LOX
UDI-Device Identifier08717648152160
UDI-Public08717648152160
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2022
Device Model Number1012455-20
Device Catalogue Number1012455-20
Device Lot Number90422G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/17/2019
Initial Date Received by Manufacturer 09/13/2019
Supplement Date Received by Manufacturer03/02/2020
Initial Report FDA Received Date10/03/2019
Supplement Report FDA Received Date03/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-