• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TISSUE SCIENCE LABORATORIES PERMACOL; MESH, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TISSUE SCIENCE LABORATORIES PERMACOL; MESH, SURGICAL Back to Search Results
Model Number P101010
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Abscess (1690); Adhesion(s) (1695); Inflammation (1932); Pain (1994); Scar Tissue (2060); Scarring (2061); Hernia (2240); Injury (2348); Impaired Healing (2378); Fluid Discharge (2686); Unspecified Tissue Injury (4559)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of an incisional hernia.It was reported that after implant, the patient experienced bowel adherent to mesh, recurrence, dense adhesions of the small bowel to the prior piece of mesh.Post-operative patient treatment included repair of recurrent incisional hernia with new bard phasix mesh and other meshes on later dates, excision of prior mesh and lysis of adhesions.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of an incisional hernia.It was reported that after implant, the patient experienced recurrence, pain, and dense adhesions of the small bowel to the prior piece of mesh.Post-operative patient treatment included repair of recurrent incisional hernia with new bard phasix mesh and other meshes on later dates, excision of prior mesh and lysis of adhesions.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of an incisional hernia.It was reported that after implant, the patient experienced inflammation, scarring, abscess, wound with associated fluid at skin surface, recurrence, pain, and dense adhesions of the small bowel to the prior piece of mesh.Post-operative patient treatment included ct scan, repair of recurrent incisional hernia with new bard phasix mesh and other meshes on later dates, excision of prior mesh and lysis of adhesions.
 
Additional Manufacturer Narrative
Correction: h6 (added patient code).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional info: a5b, d4 (unique identifier (udi) #), g1 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERMACOL
Common Device Name
MESH, SURGICAL
Manufacturer (Section D)
TISSUE SCIENCE LABORATORIES
victoria house, victoria road
aldershot, hampshire GU11 1EJ
GB  GU11 1EJ
Manufacturer (Section G)
TISSUE SCIENCE LABORATORIES
victoria house, victoria road
aldershot, hampshire GU11 1EJ
GB   GU11 1EJ
Manufacturer Contact
mike bocchino
60 middletown ave
8200 coral sea st ne
mounds view,mn, CT 55112-4391
8006464633
MDR Report Key9152937
Report Number9617613-2019-00166
Device Sequence Number2694669
Product Code FTM
UDI-Device Identifier10884523000108
UDI-Public(01)10884523000108(17)150530(10)11B1506
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K120605
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2016
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date (Section B) 02/20/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/30/2015
Device Model NumberP101010
Device Catalogue NumberP101010
Device Lot Number11B1506
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/11/2019
Supplement Date Received by Manufacturer09/11/2019
10/14/2019
06/19/2024
07/03/2024
02/11/2026
Initial Report FDA Received Date10/03/2019
Supplement Report FDA Received Date10/07/2019
11/07/2019
06/24/2024
07/09/2024
02/20/2026
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexMale
Patient Weight118 KG
Patient RaceWhite
-
-