• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION NC QUANTUM APEX; CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS Back to Search Results
Model Number 6531
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Date of Event 08/30/2019
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that shaft break occurred.While the device was outside the patient during a procedure involving a 12mm x 3.50mm nc quantum apex balloon catheter, it was noted that the device was split in its distal part.The procedure was completed with another of same device.
 
Event or Problem Description
It was reported that shaft break occurred.While the device was outside the patient during a procedure involving a 12mm x 3.50mm nc quantum apex balloon catheter, it was noted that the device was split in its distal part.The procedure was completed with another of same device.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: returned product consisted of the nc quantum apex balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There was contrast in the inflation lumen.The balloon was tightly folded.Inspection of the device revealed that the shaft was kinked at the exit notch; however, there was no breaks or fractures in the shaft or to the device.There was a kink in the shaft.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC QUANTUM APEX
Common Device Name
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key9153569
Report Number2134265-2019-11995
Device Sequence Number16817971
Product Code LOX
UDI-Device Identifier08714729783282
UDI-Public08714729783282
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K121667
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number6531
Device Catalogue Number6531
Device Lot Number0023874354
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2019
Initial Date Received by Manufacturer 09/16/2019
Supplement Date Received by Manufacturer11/20/2019
Initial Report FDA Received Date10/03/2019
Supplement Report FDA Received Date12/05/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-