| Model Number |
6531 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
08/30/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that shaft break occurred.While the device was outside the patient during a procedure involving a 12mm x 3.50mm nc quantum apex balloon catheter, it was noted that the device was split in its distal part.The procedure was completed with another of same device.
|
| |
|
Event or Problem Description
|
|
It was reported that shaft break occurred.While the device was outside the patient during a procedure involving a 12mm x 3.50mm nc quantum apex balloon catheter, it was noted that the device was split in its distal part.The procedure was completed with another of same device.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: returned product consisted of the nc quantum apex balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There was contrast in the inflation lumen.The balloon was tightly folded.Inspection of the device revealed that the shaft was kinked at the exit notch; however, there was no breaks or fractures in the shaft or to the device.There was a kink in the shaft.
|
| |
|