• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY CAPTURA PRO¿ BIOPSY FORCEPS WITH SPIKE; PTS ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED, EXEMPT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY CAPTURA PRO¿ BIOPSY FORCEPS WITH SPIKE; PTS ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED, EXEMPT Back to Search Results
Model Number G50697
Medical Device Problem Code Dull, Blunt (2407)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); No Consequences Or Impact To Patient (2199)
Date of Event 09/09/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Information regarding occupation: non-healthcare professional.Evaluation conclusion: a product evaluation was not performed because the product said to be involved was not provided to cook for evaluation.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Evaluation investigation: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.The size of tissue sample or biopsy to be excised can be controlled by how the user handles the forceps.If excessive pressure was applied during advancement into the tissue or during handle manipulation, this could have contributed to the reported observation.The instructions for use (ifu) direct the user to, ¿advance forceps to biopsy or retrieval site, then open cups and advance into tissue to be biopsied or objected to be retrieved.Using slight pressure on handle, close forceps around tissue or object.Maintain gentle handle pressure to keep cups closed and gently withdraw forceps from site.¿ the ifu warns, "these single-use biopsy forceps should only be used to biopsy tissue where possible bleeding or hemorrhage will not present a danger for patients.Adequate plans for management of potential bleeding or hemorrhage and appropriate airway management should be in place." the ifu lists potential complications as, "those associated with gastrointestinal endoscopy include but are not limited to: perforation, bleeding or hemorrhage, aspiration, fever, infection, allergic reaction to medication, hypotension, respiratory depression or arrest, cardiac arrhythmia or arrest." prior to distribution, all captura pro biopsy forceps are subjected to a visual inspection and functional test to ensure proper workability.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event or Problem Description
During an esophagogastroduodenoscopy (egd) procedure in which qty.1 of the captura pro biopsy forceps with spike, g50697, was used.When the biopsy piece of mucosa was extracted it caused approximately a 3-4 cm laceration.The laceration bled very little bit but required no intervention as it took care of itself.But the physician did intervene and placed a hemoclip just in case.The physician mentioned that the patient was in declined health which may have compromised the tissue integrity.The following was received on 03 october 2019: the customer stated that the bleed took care of it¿s self.But he did place an instinct clip just in case, to reduce risk re-bleed.A section of the device did not remain inside the patient¿s body.The patient required a clip to be placed due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTURA PRO¿ BIOPSY FORCEPS WITH SPIKE
Common Device Name
PTS ENDOSCOPIC GRASPING/CUTTING INSTRUMENT, NON-POWERED, EXEMPT
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
MDR Report Key9153826
Report Number1037905-2019-00577
Device Sequence Number3123306
Product Code PTS
UDI-Device Identifier00827002506978
UDI-Public(01)00827002506978(17)220624(10)W4229969
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 09/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/24/2022
Device Model NumberG50697
Device Catalogue NumberBF-2.4SL-230SP-40
Device Lot NumberW4229969
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/09/2019
Initial Report FDA Received Date10/03/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-