• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN MFG DC BOULDER RAPIDVAC; APPARATUS, EXHAUST, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN MFG DC BOULDER RAPIDVAC; APPARATUS, EXHAUST, SURGICAL Back to Search Results
Model Number SE3695
Medical Device Problem Codes Overheating of Device (1437); Smoking (1585); Adverse Event Without Identified Device or Use Problem (2993); Noise, Audible (3273); Intermittent Energy Output (4025)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 09/11/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, intra-operatively, the device overheated while the engine was moving.After, a loud noise was heard, the device went out and there was a lot of smoke came out of the device.An electric fire extinguisher was used on the machine.The patient inhaled the smoke.No further information available.
 
Additional Manufacturer Narrative
Evaluation summary: one device was received as a service request and a visual inspection and functional test were performed.The returned device did not meet specification as received.According to the reporter, intra-operatively, the device overheated while the engine was moving.After, a loud noise was heard, the device went out and there was a lot of smoke came out of the device.The reported condition was confirmed.Inspection of the device resulted in finding the engine blades had become loose.The investigation found the most probable cause of the reported event to be wear and tear.The motor was replaced to address the issue.The system passed all required tests.The filter check, the flow control, the footswitch operation, the interlink function, the turbo mode, and the low frequency electrical safety test passed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDVAC
Common Device Name
APPARATUS, EXHAUST, SURGICAL
Manufacturer (Section D)
COVIDIEN MFG DC BOULDER
5920 longbow drive
boulder CO 80301 3299
MDR Report Key9154599
Report Number1717344-2019-01256
Device Sequence Number1801208
Product Code FYD
UDI-Device Identifier10884524001753
UDI-Public10884524001753
Combination Product (Y/N)N
Initial Reporter CountryFI
PMA/510(K) Number
K980915
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,user facility
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSE3695
Device Catalogue NumberSE3695
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/13/2019
Initial Date Received by Manufacturer 09/11/2019
Supplement Date Received by Manufacturer11/08/2019
Initial Report FDA Received Date10/04/2019
Supplement Report FDA Received Date11/26/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-