| Model Number |
00711811 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/27/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Us endoscopy requested the device subject of the reported event be returned for evaluation, which confirmed the report of non-functional locking mechanism.The device history record was reviewed and found no noted nonconformities related to the luer lock mechanism.There have been no other complaints associated with this lot.No additional issues have been reported.
|
| |
|
Event or Problem Description
|
|
Us endoscopy received user facility medwatch #(b)(4) reporting that during a patient procedure the locking feature on their carr-lock injection needle could not be engaged.The function of the locking feature is to maintain the needle in an extended position.The user manually extended the needle and was able to successfully complete the procedure with the same device.There was no report of harm to the patient or user.
|
| |
|