Catalog Number 6051-0830S
Medical Device Problem Codes Degraded (1153); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Inflammation (1932); Reaction (2414); Metal Related Pathology (4530)
Date of Event 05/30/2012
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.The information in this report was provided by stryker orthopaedics legal affairs department.No additional information is available at this time due to the ongoing litigation.Should additional information become available, the evaluation summary will be submitted in a supplemental report.Device not returned.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his right hip on or about (b)(6) 2012.It is further alleged that he suffered injuries as a result of implantation of the device at issue, device recall and excessive levels of chromium and cobalt in his blood.Patient has not yet scheduled a surgery for explantation of the femoral head at issue.
 
Additional Manufacturer Narrative
Reported event: an event regarding abnormal ion level involving an security stem stem was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device remained implanted.Clinician review: no medical records were received for review with a clinical consultant.-product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as return of the device, pathology reports, pre- and post-operative x-rays and the revision operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If devices and/or additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
It was reported through the filing of a lawsuit that allegedly the patient was implanted with an lfit anatomic cocr v40 femoral head on his right hip on or about (b)(6) 2012.It is further alleged that he suffered injuries as a result of implantation of the device at issue, device recall and excessive levels of chromium and cobalt in his blood.Patient has not yet scheduled a surgery for explantation of the femoral head at issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SECUR-FIT MAX 132 HIP STEM NO 8
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
anna ryan
raheen business park
limerick, NJ NA
EI   NA
2018315000
MDR Report Key9156023
Report Number0002249697-2019-03397
Device Sequence Number2449238
Product Code LZO
UDI-Device Identifier07613327021028
UDI-Public07613327021028
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K051738
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2012
Reporter Type Manufacturer
Report Source Other,Consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date05/31/2017
Device Catalogue Number6051-0830S
Device Lot NumberMLEENV
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/10/2019
Supplement Date Received by Manufacturer08/21/2024
Initial Report FDA Received Date10/04/2019
Supplement Report FDA Received Date09/16/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/01/2012
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention; Hospitalization;
Patient SexMale
-
-