MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 8637-40 |
Device Problem
Difficult to Advance (2920)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 09/30/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: 8780, serial#: (b)(4), product type: catheter.Other relevant device(s) are: product id: 8780, serial/lot #: (b)(4), ubd: 01-jul-2021, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a healthcare professional via a manufacturer's representative (rep) regarding a patient who was receiving 18.3mg/ml morphine at 3.6mg/day and 22.5mg/ml bupivacaine at 4.4mg/day via an implantable infusion pump for an unknown indication for use.It was reported that during a pump and catheter replacement the healthcare professional had a hard time advancing the catheter into the spine.It was reported that the 8870 catheter was faulty and not threading.The catheter threading was attempted multiple times and it was reported that the patient had a difficult spine to access.A new 8782 catheter was used which worked fine.The issue was resolved at the time of the report, and the patient's status was noted as "alive-no injury." no further complications were reported.
|
|
Manufacturer Narrative
|
Analysis of the implantable intrathecal catheter s/n (b)(4) found that damage occurred to the guidewire during the implant procedure.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|