| Model Number |
TMTWB11 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
08/13/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Patient weight is not provided / unknown.Additional 510(k) numbers are k113753 and k112160.Device not returned to manufacturer.
|
| |
|
Event or Problem Description
|
|
Doctor reported that when viewing the x-ray after 1 month of placement of implant tmtwb11, doctor decided to change the implant to a smaller length.Tooth site # 4.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted to relay additional information.The following sections are being reported: method code was updated to 3331 and 4109.Results code was updated to 213.Conclusions code was updated to 4310.The reported event is no confirmed with the information provided and after review of the returned product.Visual inspection of the as returned product identified wear and debris functional testing to recreate the reported event could not be performed due to the nature of the device and event.A device history record (dhr) review was completed for the reported device lot.It was confirmed that all operations and inspections were executed as per applicable procedure.No deviations or non-conformance, which could have caused or contributed to the reported event was noted as part of the dhr.A complaint history review was completed for the reported device lot for similar event and no other complaint was identified.A definitive root cause for the reported event could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
| |
|
Event or Problem Description
|
|
No additional or corrected information to report.
|
| |
|