MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GNS II BICONVEX PAT 23MM; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GNS II BICONVEX PAT 23MM; PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY Back to Search Results
Catalog Number 71420566
Device Problem Break (1069)
Patient Problems No Consequences Or Impact To Patient (2199); Device Embedded In Tissue or Plaque (3165)
Event Date 09/03/2019
Event Type  malfunction  
Event Description
It was reported that after the implantation, a foreign particle was found, and it was picked up from the patient body.It was unknown if the particle was caused by the instruments or implants.
 
Manufacturer Narrative
The associated complaint devices were not returned.The particle associated with this complaint was returned and evaluated.The material analysis concluded that the metallic particle was a titanium alloy, which is consistent with the material used to manufacture the tibial base.The origin of the particle could not be confirmed.A review of complaint history did not reveal additional complaints for the listed batches.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.Without the actual devices involved our investigation cannot proceed.If the devices or new information is received in the future, the complaints can be re-opened.No further actions are being taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GNS II BICONVEX PAT 23MM
Type of Device
PROSTH, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONST, CEMENTED, POLY/METAL/POLY
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9158544
MDR Text Key161768729
Report Number1020279-2019-03557
Device Sequence Number1
Product Code JWH
UDI-Device Identifier03596010208255
UDI-Public03596010208255
Combination Product (y/n)N
PMA/PMN Number
K951987
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 12/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/04/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number71420566
Device Lot Number19EM01705
Was Device Available for Evaluation? No
Date Manufacturer Received12/04/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-