Model Number MMT-551LNAH
Medical Device Problem Codes Obstruction of Flow (2423); No Apparent Adverse Event (3189)
Health Effect - Clinical Code Hyperglycemia (1905)
Date of Event 10/02/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Customer reported via phone call that they experienced high blood glucose level.The customer¿s blood glucose level was 426 mg/dl at the time of the incident.Customer stated insulin pump had no delivery alarm and insulin was exited.Troubleshooting was performed.Customer was advised that the insulin pump needed to be replaced and to revert to the backup plan per health care professional instructions.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
No unexpected no delivery alarm noted during testing.Device passed rewind test, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and displacement test.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-551LNAH
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key9159949
Report Number2032227-2019-74380
Device Sequence Number2449314
Product Code OZO
UDI-Device Identifier00643169933002
UDI-Public(01)00643169933002
Combination Product (Y/N)N
Initial Reporter StateWV
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/25/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberMMT-551LNAH
Device Catalogue NumberMMT-551LNAH
Device Lot NumberA5551LNAHJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/17/2019
Initial Date Received by Manufacturer 10/02/2019
Supplement Date Received by Manufacturer10/24/2019
Initial Report FDA Received Date10/05/2019
Supplement Report FDA Received Date10/25/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRN-UNK-RSVR,UNOMED INF SET
Outcome Attributed to Adverse Event Other;
Patient Age61 YR
Patient Weight127
-
-