| Model Number |
763662 |
| Medical Device Problem Codes |
Crack (1135); Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/27/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The incident sample has been requested but to date has not been received for evaluation. if the sample is received, or if additional information pertinent to the incident is obtained a follow-up report will be submitted. as part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.
|
| |
|
Event or Problem Description
|
|
The customer reported that the feeding bag had many slits and there were leaks near the tubing.The feed flowed freely out of the bag and onto the floor.
|
| |
|
Additional Manufacturer Narrative
|
|
The device history record review of the reported lot number shows evidence that the product was released according to all established procedures and qa documentation.One used sample was received for evaluation.A visual inspection was performed, and it was detected that the bag has a cut near the tubing connection.The condition was not determined as a manufacturing issue since after the production line assessment no potential root causes were detected and there is no trend for a damaged bag related to this part number.It is possible that the condition occurred post manufacturing during use.This complaint will be closed with no further action and will be used for tracking and trending purposes.
|
| |
|