| Model Number |
109.781 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Pain (1994); Tissue Damage (2104)
|
| Date of Event |
06/02/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The information provided in this report was based on information given
by the dentist in neodent¿s products warranty form that was recorded in
the system and is used by both the manufacturer and the subsidiary.The
original complaint and the product were received by the subsidiary and
did not arrive in the manufacturer yet.When this happens, a new
evaluation of the complaint will be carried out and, if necessary, a new
follow-up report will be send.
|
| |
|
Event or Problem Description
|
|
(b)(4)- the dentist reported that 7 months and 9 days after the dental implant was installed in intraoral region 30#, its fracture was observed.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional evaluation c.Conclusion: 18/failure to follow instructions additional evaluation c.Conclusion: 50/ device failure related to patient's cond.The complaint was received by manufacturer and, after analysis, new informations were obtained.A follow-up is being sent to correct these informations according to the evaluation made.The investigation of the complaint was carried out considering the analysis of the information given by the dentist in the guarantee form, visual conference of the product returned by the dentist and the evaluation of relevant production records.Considering the surgical methodology, it was seen that the dentist has used sequence of drills different than the one recommended for the implant installation.
|
| |
|
Event or Problem Description
|
|
(b)(4) - the dentist reported that 7 months and 9 days after the dental implant was installed in intraoral region 30#, its fracture was observed.
|
| |
|