| Catalog Number |
2N3342 |
| Medical Device Problem Codes |
Fluid/Blood Leak (1250); Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/05/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the tubing of a non-dehp three lead extension set had a hole which caused a leak.It was further reported that the hole was located between the upper two branches of tubing and the lower branch, in a portion of the tubing with no slide clamp.The issue was identified during use on a baby in the nicu (neonatal intensive care unit).It was reported that the patient was already on unspecified antibiotics at the time of the event.The set was in use to deliver total parental nutrition (tpn) to the patient.There was no patient injury or medical intervention associated with this event.No additional information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information : the device was received for evaluation.A visual inspection was performed using the naked eye which observed that all components are correctly placed and according to product specifications, however, a pin hole was observed in the tubing.Due to the nature of the returned sample no functional testing was performed.The reported problem was verified.The cause of the condition was not determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
| |
|
Search Alerts/Recalls
|