| Model Number |
26924 |
| Medical Device Problem Codes |
Difficult to Insert (1316); Premature Activation (1484)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
09/16/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that the stent prematurely deployed.A 6x80x130 innova self expanding stent was selected for use for a chronic total occlusion (cto) of the superficial femoral artery (sfa) stenting procedure.The innova delivery system was advanced over a non-bsc guidewire inside a non-bsc sheath to the target lesion.There were difficulties encountered entering the sheath.The physician realized on fluoro that something was not right before the physician attempted deploy the stent.The catheter and stent were removed from the patient and it was noted stent was partially prematurely deployed without expansion.The procedure was completed with another innova stent without any issues.There were no patient complications reported.
|
| |
|