STRYKER ORTHOPAEDICS-MAHWAH HMRS/G I-II. CIRCLIPS; PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 6466-9-230
Medical Device Problem Codes Component Incompatible (1108); Inadequacy of Device Shape and/or Size (1583)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
Date of Event 09/03/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A supplemental report will be submitted upon completion of the investigation.
 
Event or Problem Description
It was reported that the circlip used to lock the axle in place was too small to properly lock the axle.Update: 5 minutes surgical delay.
 
Event or Problem Description
It was reported that the circlip used to lock the axle in place was too small to properly lock the axle.Update: 5 minutes surgical delay.
 
Additional Manufacturer Narrative
An event regarding assembly issue involving a custom hmrs circlip was reported.The event was not confirmed.Method & results: -device evaluation and results: visual inspection was performed by lisi medical and indicated the following comments: the hmrs/g i-ii.Circlips showed distortion: the device is oval and not circular dimensional inspection was performed by lisi medical and indicated the following comment: dimensional inspection was performed according to the drawing number 454-b revision c: -wire diameter 0.90+/-0.02 mm: 0.88 mm - compliant -height 13.00 +/- 0.215 mm: 12.92 mm - compliant -flatness (visual inspection) - compliant -gauge 6.00+/-0.15 mm: 5.35 mm - not compliant the gauge of the device is dimensionally not compliant -clinician review: no medical records were received for review with a clinical consultant.-device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the reported lot.Conclusion the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports and return of the mating devices are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HMRS/G I-II. CIRCLIPS
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key9162681
Report Number0002249697-2019-03415
Device Sequence Number13931399
Product Code KRO
UDI-Device Identifier04546540276520
UDI-Public04546540276520
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K122015
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2022
Device Model Number6466-9-230
Device Catalogue Number6466-9-230
Device Lot NumberG7265432
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/24/2019
Initial Date Received by Manufacturer 09/09/2019
Supplement Date Received by Manufacturer12/23/2019
Initial Report FDA Received Date10/07/2019
Supplement Report FDA Received Date01/21/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-