| Model Number |
6466-9-230 |
| Medical Device Problem Codes |
Component Incompatible (1108); Inadequacy of Device Shape and/or Size (1583)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/03/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A supplemental report will be submitted upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
It was reported that the circlip used to lock the axle in place was too small to properly lock the axle.Update: 5 minutes surgical delay.
|
| |
|
Event or Problem Description
|
|
It was reported that the circlip used to lock the axle in place was too small to properly lock the axle.Update: 5 minutes surgical delay.
|
| |
|
Additional Manufacturer Narrative
|
|
An event regarding assembly issue involving a custom hmrs circlip was reported.The event was not confirmed.Method & results: -device evaluation and results: visual inspection was performed by lisi medical and indicated the following comments: the hmrs/g i-ii.Circlips showed distortion: the device is oval and not circular dimensional inspection was performed by lisi medical and indicated the following comment: dimensional inspection was performed according to the drawing number 454-b revision c: -wire diameter 0.90+/-0.02 mm: 0.88 mm - compliant -height 13.00 +/- 0.215 mm: 12.92 mm - compliant -flatness (visual inspection) - compliant -gauge 6.00+/-0.15 mm: 5.35 mm - not compliant the gauge of the device is dimensionally not compliant -clinician review: no medical records were received for review with a clinical consultant.-device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.-complaint history review: there have been no other similar events for the reported lot.Conclusion the exact cause of the event could not be determined because insufficient information was provided.Additional information including operative reports and return of the mating devices are needed to fully investigate the event.If further information becomes available or the product is returned, this investigation will be re-opened.
|
| |
|
Search Alerts/Recalls
|