STRYKER ORTHOPAEDICS-MAHWAH TRIATHLON CR X3 TIBIAL INSERT; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO Back to Search Results
Catalog Number 5530-G-409
Medical Device Problem Codes Failure to Osseointegrate (1863); Osseointegration Problem (3003)
Health Effect - Clinical Codes Injury (2348); Ambulation or Postural Difficulties (2544); Inadequate Osseointegration (2646)
Date of Event 09/10/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.If additional information is received, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Revision aseptic loosening.
 
Additional Manufacturer Narrative
An event regarding loosening involving a triathlon insert was reported.Conclusion: a full complaint investigation was completed for both the femoral component and the baseplate within this pi.Loosening refers to loss of fixation between bone and implant/bone cement or failure to fixate.The insert has no direct contact with the patients bone.The insert is an integral part of the tibial component construct and has to be revised as the baseplate it was locked into was explanted.Based on the information provided there is no indication or allegation that device reported in this investigation may have contributed to the event.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
Revision aseptic loosening.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIATHLON CR X3 TIBIAL INSERT
Common Device Name
PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key9163184
Report Number0002249697-2019-03417
Device Sequence Number1776665
Product Code MBH
UDI-Device Identifier07613327050479
UDI-Public07613327050479
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K141056
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2021
Device Catalogue Number5530-G-409
Device Lot Number3V6H3H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/23/2019
Initial Date Received by Manufacturer 09/10/2019
Supplement Date Received by Manufacturer12/19/2019
Initial Report FDA Received Date10/07/2019
Supplement Report FDA Received Date01/17/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age65 YR
Patient Weight103
-
-