The returned device was patent.The valve did not pass siphon, leakage, pressure flow characteristics and preimplantation testing.Though no tears were evident during visual inspection, air bubbles were observed during leakage testing, which indicates possible tear on the delta chamber and outlet connector junction.Ifu states that ¿use of sharp instruments while handling these devices can nick or cut the silicone elastomer, resulting in leakage and necessitating shunt revision.Care must be taken when closing incisions to ensure that the valves are not cut or nicked by suturing needles.¿ review of dhr for lot # e45664did not identify any finding correlated to the complaint.All valves are 100% tested at the time of manufacture.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|