|
It was reported that, during a rotator cuff repair surgery, a straight probe was found to be cracked in the patient articulation.The cracked tips were removed by using a punch basket.A back-up device was available to complete the procedure in the scheduled time.The patient was not harmed.
|
|
One 010001 straight probe used for treatment, and not returned for evaluation.The complaint stated: ¿it was found to be cracked in the patient articulation.¿ due to unavailability, conclusions, investigation and evaluation were limited.If objective evidence becomes available to assist with evaluation, the complaint will be revisited.Factors affecting device performance may include: device storage, device ability, surgical ability and surgical site preparation according to instructions for use.Instructions for use confirms instructions, precautionary statements and recommendations for proper use of product.Influences that could compromise product performance or integrity include: compliant use of instrument.Careful cleaning and sterilization of product.Entanglement with other instruments or devices during use or cleaning.Pressure or excess torque applied.Per instructions for use: ¿these instruments are designed for repeated use with proper care and handling.¿ instrument tips may inadvertently be twisted or caught on other instruments or the basket during cleaning.This can lead to compromised strength.¿as with any surgical instrument, careful attention should be made to assure that excessive force is not placed on the instrument.Excessive forces applied to the instrument can result in the instrument¿s failure.Do not use these instruments as levers for manipulating hard tissue or bone.Excessive force should not be applied to the instrument when manipulating soft tissue, bone, or hard objects.Misuse of these instruments may result in bent distal tips or jaws; and dull or uneven cutting edges.¿ product met specifications upon release to distribution.
|