|
The concerto pushwire was found broken in three locations.The first break was at the positive load indicator with the release wire was also found broken.The second break was found between the positive load indicator and break indicator and retained by the release wire.The third break was found distal to the break indicator at the coupler tube/hypotube junction also retained by the release wire.This is indicative that manual detachment was attempted at these locations.Bends and kinks were found between the positive load indictor and break indicator, and at ~30.2cm, ~36.8cm, and ~141.4cm from the proximal end of the most distal segment.The coin was found pulled back from the lumen stop and the shield coil was present but damaged.The implant coil was already detached and not returned for analysis.All other subassemblies appeared to be normal and no other anomalies were observed.Based on the device analysis and reported information, the customer¿s report of ¿coil resistance/stuck in catheter¿ could not be confirmed and the root cause could not be determined.The concerto coil could not be used for resistance testing with an in-house micro catheter as the implant coil was already detached.There was evidence of detachment attempts, and the coin was pulled back from the lumen stop, releasing the implant coil as intended.Bends and kinks were found throughout the length of the pushwire, however, the cause could not be determined.Potential causes are user advancing the coil against resistance or damaged during return shipping to medtronic for analysis as the device was returned without its protective dispenser coil and introducer sheath.The shield coil was found damaged; however, the cause could not be determined.It is possible the shield coil became damaged when attempted to advance the coil against resistance or during return shipping to medtronic.Since the micro catheter used in the event was not returned, and the model or lot number was not provided, any contribution of the micro catheter towards the reported ¿coil resistance/stuck in catheter¿ could not be assessed.There was no indication that the event was related to a manufacturing issue, therefore a device history review is not required.If information is provided in the future, a supplemental report will be issued.
|