Model Number G21396
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/09/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
1 x zebd-7-7 of lot # c1572371 was returned to cirl for evaluation open in its original packaging.The device involved in the complaint was evaluated in the laboratory on 27th september 2019.A kink was observed on the 5th porthole at the tapered end.A 0.035-inch wireguide would not pass the kink.Prior to distribution all zebd-7-7 are subjected to visual and functional checks to ensure device integrity.These inspections and functional checks are outlined in internal procedures in place in cirl.A review of the manufacturing records for zebd-7-7 of lot number c1572371 did not reveal any discrepancies that could have contributed to this complaint issue.The review of relevant manufacturing records, confirms the failure mode has previously occurred with the current lot number.Based on the information available to date, there is no evidence to suggest that there are any manufacturing issues associated with lot number c1572371 it should be noted that the instructions for use (ifu0045-7) states the following: ¿visually inspect with particular attention to kinks, bends and breaks.If any abnormality is detected that would prohibit proper working condition, do not use.¿ ¿care must be exercised when straightening the pigtail curls in order to avoid kinking or breaking the stent.¿ a definitive root cause could not be determined as the circumstances of use cannot be replicated in the laboratory.A possible root cause could be attributed to the use of a non-recommended wire guide with the device.It may be noted the device label indicates a 0.035¿ wire guide for use with this device and all simulated use testing to date has been completed using a 0.035¿ wire guide.As the smaller diameter 0.025¿ wire guide occupies less space in the lumen of the pigtail stent it is likely that the extent of the curvature of the pigtail will be greater when a 0.025¿ wire guide is used.As per additional information received a 0.025" wire guide was used.Complaint is confirmed as the failure was verified in the laboratory.According to the initial reporter, the patient did not experience any adverse effects.Complaints of this nature will continue to be monitored for potential emerging trends.
 
Event or Problem Description
Event is fda mdr reportable based on the device malfunction reporting precedence for this device family for the issue of ¿stent kinked/ bent'.After placing olympus's wire guide visiglide 0.025inch in the common bile duct, the user advanced the zebd- stent over the wire guide.However, the wire guide could not pass through the stent since the pigtail became kinked.The stent was replaced with another manufacturer's erbd stent and the procedure was completed.There have been no adverse effects to the patient reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMON BILIARY STENT SET
Common Device Name
FGE CATHETER, BILIARY, DIAGNOSTIC
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
MDR Report Key9165523
Report Number3001845648-2019-00513
Device Sequence Number16834862
Product Code FGE
UDI-Device Identifier00827002213968
UDI-Public(01)00827002213968(17)220102(10)C1572371
Combination Product (Y/N)N
PMA/510(K) Number
K851962
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 09/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/02/2022
Device Model NumberG21396
Device Catalogue NumberZEBD-7-7
Device Lot NumberC1572371
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/24/2019
Was the Report Sent to FDA? No
Distributor Facility Aware Date09/09/2019
Event Location Hospital
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/10/2019
Initial Report FDA Received Date10/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-