STRYKER ORTHOPAEDICS-MAHWAH 28MM STD LFIT V40 HEAD; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Catalog Number 6260-9-128
Medical Device Problem Codes Degraded (1153); Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Inflammation (1932); Injury (2348); Reaction (2414)
Date of Event 09/10/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.Device not returned.
 
Event or Problem Description
Patient presented with a dislocated hip and was scheduled for revision surgery on (b)(6) 2019.Upon exposure it was noted that the poly liner had disassociated from the femoral head of the mdm construct.Dr.Mentioned prior to the surgery that the patient had elevated cobalt ion levels.The femoral head was removed and replaced with a c-taper adaptor sleeve and a +7.5 delta ceramic head.The cocr cementless liner was removed and replaced with an x 3 0 degree liner.There was corrosion on the top of the cementless liner.Update: left side.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
28MM STD LFIT V40 HEAD
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key9165811
Report Number0002249697-2019-03450
Device Sequence Number13994429
Product Code JDI
UDI-Device Identifier07613327013023
UDI-Public07613327013023
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K173499
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2019
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date10/31/2016
Device Catalogue Number6260-9-128
Device Lot Number38087002
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/10/2019
Initial Report FDA Received Date10/08/2019
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age55 YR
-
-