| Model Number |
4252520-02 |
| Medical Device Problem Code |
Contamination /Decontamination Problem (2895)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/12/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical inc.Internal report number (b)(4).The investigation into this reported event is ongoing.A follow-up report will be submitted when the results of the investigation are available.
|
| |
|
Event or Problem Description
|
|
As reported by user facility: small worm or larvae in catheter.
|
| |
|
Additional Manufacturer Narrative
|
|
This report has been identified as b.Braun medical inc.Internal report number (b)(4).The actual device involved in the reported incident was not returned, however a photo was provided for evaluation.A small particle could be seen in the photo, however without the sample to evaluate no analysis could be performed to identify the particle material.Without the actual sample a thorough sample analysis could not be performed and no specific conclusions can be drawn.Review of the device history record performed for the reported lot number did not reveal any abnormalities or non-conformances of this nature.If additional pertinent information becomes available a follow-up report will be filed.
|
| |
|
Search Alerts/Recalls
|