• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK ENDOSCOPY FUSION® PRE-LOADED WITH ACROBAT WIRE GUIDE; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK ENDOSCOPY FUSION® PRE-LOADED WITH ACROBAT WIRE GUIDE; KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES) Back to Search Results
Model Number G34693
Device Problem Material Separation (1562)
Patient Problems No Consequences Or Impact To Patient (2199); Foreign Body In Patient (2687)
Event Date 09/12/2019
Event Type  malfunction  
Manufacturer Narrative
Occupation: non-healthcare professional investigation evaluation: our laboratory evaluation of the product said to be involved confirmed and determined the cutting wire securing component located near the distal end of the sphincterotome has disconnected from the catheter.The condition of the device prohibited a functional test of the used device because the cutting wire securing component located near the distal end of the sphincterotome has disconnected from the catheter.The cutting wire is intact and remains securely attached to the sphincterotome at the proximal end.However, due to the catheter and securing component disconnection, the distal end of the cutting wire is no longer connected to the sphincterotome catheter.A section of the cutting wire securing component has broken and detached from the device.The broken section is estimated to be 3mm in length and 0.5mm in diameter.The broken section was not included in the return.The cutting wire shows evidence of cautery application.A product discrepancy or anomaly that could have contributed to this reported occurrence was not observed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a definitive cause for this observation could not be determined because the actual use conditions could not be duplicated in the laboratory setting.Due to a variety of clinical conditions such as patient anatomy, endoscope position or progression of disease state, we could not reproduce the actual conditions of product usage during our laboratory analysis.This limits our ability to conclusively determine a cause.Separation of the cutting wire securing component and the catheter can occur if the tip of the sphincterotome is over flexed.The instructions for use caution the user: "do not over flex or bow tip beyond 90 degrees, as this may damage or cause cutting wire to break." other factors that can contribute to separation of the cutting wire securing component and the catheter include manipulating the handle with the catheter in a coiled position or with the precurved stylet inside the cannulating tip.The instructions for use advise the user: "upon removing device from package, uncoil and straighten sphincterotome.Carefully remove precurved stylet from cannulating tip." the instructions for use contain the following comment: "note: do not exercise handle while device is coiled or precurved stylet is in place, as this may cause damage to sphincterotome and render it inoperable." prior to distribution, all fusion pre-loaded with acrobat wire guide are subjected to a visual inspection and functional test to ensure device integrity.The functional test includes bowing the sphincterotome to ensure the distal end responds to handle manipulation.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a review of the complaint history was conducted.The likelihood of occurrence is considered rare.Corrective action is not warranted at this time based on the quality engineering risk assessment.Quality assurance will continue to monitor for complaint trends and reassess the risk assessment results as post market feedback continues to become available.
 
Event Description
During an endoscopic retrograde cholangiopancreatography (ercp), the physician used a cook fusion® pre-loaded with acrobat wire guide sphincterotome.The cutting wire detached from sphincterotome during operation.Our evaluation of the returned device determined a part of the device is missing and was not included in the return of the device.This information was communicated to the user facility and the location of the missing section is unknown.The initial reporter stated that a section of the device did not remain inside the patient¿s body; however the location of the missing section detected during our laboratory evaluation is unknown.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION® PRE-LOADED WITH ACROBAT WIRE GUIDE
Type of Device
KNS, UNIT, ELECTROSURGICAL, ENDOSCOPIC (WITH OR WITHOUT ACCESSORIES)
Manufacturer (Section D)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer (Section G)
COOK ENDOSCOPY
4900 bethania station rd
winston-salem NC 27105
Manufacturer Contact
scottie fariole
4900 bethania station rd
winston-salem, NC 27105
3367440157
MDR Report Key9167599
MDR Text Key219780048
Report Number1037905-2019-00586
Device Sequence Number1
Product Code KNS
UDI-Device Identifier10827002346939
UDI-Public(01)10827002346939(17)220415(10)W4203321
Combination Product (y/n)N
Reporter Country CodeGB
PMA/PMN Number
K052051
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 09/13/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/08/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/15/2022
Device Model NumberG34693
Device Catalogue NumberFS-OMNI-ACRO-35-260
Device Lot NumberW4203321
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/04/2019
Date Manufacturer Received09/13/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/18/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-