| Model Number |
7267V |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
09/19/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The outer layer of material (blue) was separated.The separation point was stretched suggesting the dilator was pulled against resistance.The sheath was also shown in the picture, but it was undeterminable if damage was present that could have led to the separation.Without the device returning further investigation could not be completed.
|
| |
|
Event or Problem Description
|
|
Lost part of the dilator in patient during procedure.Not noticed until scrub tech noticed part of the dilator was missing when she noticed it on the back table.They fluoro'd patient but couldn't see the dilator.Embolized and still in the body.Patient was treated for scheduled procedure through existing access.Later ct scan revealed it is still in the common femoral artery and doctor is determining whether to go back in after it or leave it.After multiple attempts, no further information received, however, images were received to support investigation.
|
| |
|