The reported event was unconfirmed.No balloon burst was noted in the returned lubricath catheter sample.When the catheter was inflated with 35cc of water using a leur lock syringe, the balloon was inflated and no leaks were noted after allowing the catheter to sit for over about a minute.The catheter was then deflated using a leur lock syringe with no issues.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "sterile unless package is opened or damaged.Warning: on catheter, do not use ointments or lubricants having petrolatum base.They will damage latex and may cause the balloon to burst.Caution: do not aspirate urine through drainage funnel wall.Storage: store catheters at room temperature away from direct exposure to light, preferably in the original box.Recommended inflation capacities 3cc balloon: use 5ml sterile water, 5cc balloon: use 10ml sterile water, 15cc balloon: use 20ml sterile water, 20cc balloon: use 25ml sterile water, 30cc balloon: use 35ml sterile water, 40cc balloon: use 45ml sterile water, 75cc balloon: use 80ml sterile water.Do not exceed recommended capacities.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable local, state, and federal laws and regulations.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.".
|