| Model Number |
9734403 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
08/27/2019
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No patient information provided as no patient was involved in this concern.The probe was returned to the manufacturer for evaluation.Testing found that the probe tip was twisted and that there were impact marks on the back end of the instrument.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation device being used outside of procedure.It was reported that there was a box full of damaged instrument.The manufacturing representative stated that some of the tips were visibly bent and the spine clamps would not engage.There was no patient present when this issue was identified.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a navigation device being used outside of procedure.It was reported that the device was found in a hospital box labeled ¿damaged¿ in the sterile processing department at the facility.The box contained several used instruments.There was no further context able to be provided by the initial reporter.There was no patient present when this issue was identified.
|
| |
|