According to the reporter, while transecting lung tissue during an open lobectomy, the surgeon was able to press the green button and was able to squeeze the handle, but the handle did not fire.Another handle was used to complete the case.There was no patient injury.
|
|
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.Initial visual inspection of the instrument noted that the firing knobs were retracted, and the articulation lever was in neutral position.Functionally, the instrument was loaded with a representative single use loading unit.During testing of the instruments, a skip was noted in the units firing stroke after closure of the reload jaws.An access hole was cut into the instrument body for visualization of the firing rack.This examination noted sheared teeth on the firing rack.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of the sheared teeth on the firing rack may occur in any of the following circumstances: 1.Firing over tissue that is beyond the recommended thickness range.2.Firing with an obstacle incorporated in the jaws.In either of these circumstances, it will become increasingly difficult to actuate the firing handle and the instrument return knobs will be difficult to retract.In addition, staples may not form properly, and tissue may not be fully transected.The information booklet which accompanies each product shipment offers the following as a warning and precaution."1.Preoperative radiotherapy may result in changes to tissue.These changes may, for example, cause the tissue thickness to exceed the indicated range thickness for the staple size.Careful consideration should be given to any pre-surgical treatment the patient may have undergone and in corresponding selection of staple size.The root cause of the observed damage was due to the product not being used as intended which caused or contributed to the reported condition.No further actions have been deemed necessary at this time.If information is provided in the future, a supplemental report will be issued.
|