| Catalog Number |
205547000 |
| Medical Device Problem Codes |
Device-Device Incompatibility (2919); Scratched Material (3020)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary = > examination of the returned instrument confirmed the complaint.The root cause is attributed to a combination of the damage to the trial cup and the ball bearings on the handle not functioning as intended due to poor cleaning.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.(b)(4) reports that while inspecting instruments prior to sterilization, it was noticed the size 47 head trial would not stay on the trial handle appropriately.Easy to pull off and not used in surgery and surgery was not delayed.A functional check with the returned trial cup and handle confirmed the complaint; the two mating instruments would not assemble as intended.The complaint sample consisted of (1) 205547000 s/c & mod cath trial 47/28, lot number u4bff1.Review of the as400 found this lot was placed into distribution in 2000 and is 19-years old.Visual analysis found the trial cup is significantly damaged all over including inside the perimeter where the handle assembles to the trial cup.The trial cup exhibits numerous deep gouges, nicks and scratches.The root cause is attributed to a combination of the damage to the trial cup and the ball bearings on the handle not functioning as intended.A search of the complaint database found poor, repeated cleaning over time may be a contributing factor in loss of the handle ball plunger function along with heavy usage.The investigation could not verify or identify any product contribution to the reported event with the information provided.Based on the investigation, the need for corrective action is not indicated.Complaints will be monitored under post market surveillance (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that while inspecting instruments prior to sterilization, it was noticed the size 44, 45, 46, 47, and 48 head trials would not stay on the trial handle appropriately.They are easy to pull off.These were not used in surgery and surgery was not delayed.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Additional narrative: product complaint (b)(4).Investigation summary : examination of the returned instrument confirmed the complaint.The root cause is attributed to a combination of the damage to the trial cup and the ball bearings on the handle not functioning as intended due to poor cleaning.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|