|
This report contains supplemental information for mentor psr reference number (b)(4).This device report is being submitted late as a result of exemption transition period following the revocation of mentor¿s psr exemption #e2007003.Device evaluation summary: upon initial inspection, a tear measuring approximately 15 cm located on the anterior aspect was noticed.No additional anomalies were discovered.Since the authorization for return and examination of medical device form was not signed and/or returned, mentor product analysis lab was precluded from further evaluating the device.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.A possible cause for the condition reported is due to excessive force to the chest; trauma; compression during mammographic imaging; and severe capsular contracture.The reported complaint was confirmed.It should be noted that as part of our quality process, each device is visually inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.Based on the information reported and/or the product investigation, there is no evidence that the issue is related with the manufacturing process.Reason for device explant and/or reoperation: rupture.Manufacturer's reference number: (b)(4).
|